Performance Specifications
10-20 Data Transfer
Data Transfer
Performance Specifications
For safety and environmental specifications, please refer to Chapter 11,
“Monitor Installation and Patient Safety”.
Data
Transfer
Module
Transfer Time
A typical transfer time for 24 hours of data is under 30 seconds. The
transfer time could be as high as 60 seconds for highly customized
trends.
Back up Memory Time
The Data Transfer Module will store data for a minimum of 1 hour under
the following conditions:
the module is out of the monitor,
the module is in the rack of a powered Off monitor,
the monitor has a power failure.