Changing the Patient Category
Setting up your Monitor 3-45
Setting up your
Monitor
Changing the Patient Category
The monitor processes patient signals in different ways (using different
algorithms) depending on the patient category. The patient category
within a Configuration Set can be changed during normal monitoring.
In general, all the parameter settings and patient data management
information are kept if you change the patient category without changing
Configuration Set.
Possible exceptions occur because measurement ranges and alarm limit
ranges for some parameters are different for each patient category (see
the following list). If active alarm limits are no longer within range after
you change patient category, the alarm limits are automatically adjusted
to the closest possible values within the new range.