Philips V26CT Instrukcja Użytkownika Strona 281

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 280
Viewing Neonatal Events
Neonatal Event Review 9-7
Neonatal Event
Review
Event
Threshold Line
The patient monitors can be adjusted in monitoring mode so that an event
is either triggered by an alarm or by criteria defined specifically by the
user.
The event trigger level is represented in the graph by a threshold line,
located in each graph channel. This line enables the user to see the
following:
1. Whether a trigger is active. This means that the parameter is
switched on, the parameter alarm is switched on, the main alarms
are switched on and valid numerical values are available. The
threshold line is indicated on the graph by an unbroken line.
2. At what value it has been set.
3. Whether it has been changed during the 24-hour review period.
(1) Event Threshold Line
(2) Non Active Threshold Line
(3) Trigger Level Change
Note— When the patient monitor is connected to the Philips Information
Center, the Bradycardia threshold line is displayed as a shaded area.
Przeglądanie stron 280
1 2 ... 276 277 278 279 280 281 282 283 284 285 286 ... 391 392

Komentarze do niniejszej Instrukcji

Brak uwag