PHIPhilips M1165/66/67/75/76/77ACMS Patient Monitoring System andPhilips M1205AV24 and V26 Patient MonitorUser’s Reference ManualVolume 1System Inform

Strona 2 - Important

Indications for Usex • Replacement Parts - It is highly recommended that only Philips Medical Systems recommended parts and accessories be use

Strona 3 - iii

oxyCRG Display3-26 Setting up your Monitor Setting up your Monitor Approximately the bottom two thirds of the Main Screen area are covered with

Strona 4 - Electromagnetic Interference

oxyCRG DisplaySetting up your Monitor 3-27Setting up your Monitor is switched off, and one which happens after the power is restored.• If you are u

Strona 5 - Intended Use

CSA Display (CMS only)3-28 Setting up your Monitor Setting up your Monitor CSA Display (CMS only)The CSA (Compressed Spectral Array) is a format

Strona 6 - Indications for Use

CSA Display (CMS only)Setting up your Monitor 3-29Setting up your Monitor Main ScreenNotes on CSA• The CSA information is overlayed on the realtime

Strona 7 - vii

Wave Replace3-30 Setting up your Monitor Setting up your Monitor Wave ReplaceWave Replace enables the monitor to fill a currently unused channel

Strona 8

Trace ModeSetting up your Monitor 3-31Setting up your Monitor Trace ModeTrace Mode allows you to choose between fixed and moving traces: Procedure

Strona 9 - ix

Configuring a Second Independent Display (CMS only)3-32 Setting up your Monitor Setting up your Monitor Configuring a Second Independent Display

Strona 10 - Notice to the User

Other Functions You Can ConfigureSetting up your Monitor 3-33Setting up your Monitor Other Functions You Can ConfigureParameters On/OffIndividual p

Strona 11 - Manufacturer´s Address

Other Functions You Can Configure3-34 Setting up your Monitor Setting up your Monitor 4. Press to return to the Main Screen. You can also swi

Strona 12

Adjusting the Volume ControlSetting up your Monitor 3-35Setting up your Monitor Adjusting the Volume ControlYou can adjust the QRS and alarm volume

Strona 13 - Contents

Responsibility of the Manufacturer xiResponsibility of the ManufacturerPhilips Medical Systems only considers itself responsible for any effe

Strona 14 - Contents-2

Adjusting the Volume Control3-36 Setting up your Monitor Setting up your Monitor 2. Press to return to the Selection Window or to return to t

Strona 15 - Setting up your Monitor 3-1

Adjusting the Date and TimeSetting up your Monitor 3-37Setting up your Monitor Adjusting the Date and TimeYou can set the date and time to be displ

Strona 16 - Contents-4

Adjusting the Date and Time3-38 Setting up your Monitor Setting up your Monitor 3. Press to return to the Selection Window or to return to th

Strona 17 - Recording Functions 6-1

Selecting Waves for Central RecordersSetting up your Monitor 3-39Setting up your Monitor Selecting Waves for Central RecordersYou can select two wa

Strona 18 - Contents-6

Configuring Module, Bedside and Central Recordings3-40 Setting up your Monitor Setting up your Monitor Configuring Module, Bedside and Central R

Strona 19 - Trends and Calculations 8-1

Other Patients ControlsSetting up your Monitor 3-41Setting up your Monitor Other Patients ControlsIf your system is connected via the Philips monit

Strona 20 - Data Transfer 10-1

The Status Log Function3-42 Setting up your Monitor Setting up your Monitor The Status Log FunctionThe softkey is used by service engineers. I

Strona 21 - Maintenance 13-1

The Monitor Revision FunctionSetting up your Monitor 3-43Setting up your Monitor The Monitor Revision FunctionThe softkey is used by service engi

Strona 22 - Contents-10

Changing Default Settings and Patient Category3-44 Setting up your Monitor Setting up your Monitor Changing Default Settings and Patient Categor

Strona 23 - Monitors

Changing the Patient CategorySetting up your Monitor 3-45Setting up your Monitor Changing the Patient CategoryThe monitor processes patient signals

Strona 24 - Introduction

Responsibility of the Manufacturerxii

Strona 25 - M1167/77A System

Changing the Patient Category3-46 Setting up your Monitor Setting up your Monitor The following table details the differences between the patien

Strona 26

Changing the Patient CategorySetting up your Monitor 3-47Setting up your Monitor NBP The patient's tolerance for the cuff pressure (i.e. stre

Strona 27

Changing the Patient Category3-48 Setting up your Monitor Setting up your Monitor These safety values effect the measurement range of each categ

Strona 28

Changing the Patient CategorySetting up your Monitor 3-49Setting up your Monitor 1The figure given for overpressure protection is the maximum press

Strona 29 - V24 and V26

Changing the Patient Category3-50 Setting up your Monitor Setting up your Monitor RecommendationChoose the patient category according to the QRS

Strona 30 - Control

Changing the Patient CategorySetting up your Monitor 3-51Setting up your Monitor Heart Rate (HR) / PulseThe expected range can be estimated by perf

Strona 31

Changing the Patient Category3-52 Setting up your Monitor Setting up your Monitor CO2Note—The Apnea alarm delay time cannot be changed Adult/Ped

Strona 32 - Handheld

Changing the Configuration SetSetting up your Monitor 3-53Setting up your Monitor Changing the Configuration SetIf you want to change the applicati

Strona 33 - External

Changing the Configuration Set3-54 Setting up your Monitor Setting up your Monitor • Remove the plug-in modules from the rack before starting th

Strona 34 - Hardkey

Changing Operating ModesSetting up your Monitor 3-55Setting up your Monitor Changing Operating ModesThis function will be mainly used by service en

Strona 35 - Airway Gases/Ventilation

ContentsContents-1This book is divided into three volumes.This volume contains chapters 1 to 13 (see the following table of contents for more details)

Strona 36 - V26CT/V24CT Power Supply

Changing Operating Modes3-56 Setting up your Monitor Setting up your Monitor Returning to Monitoring ModeTo exit Config, Demo or Service Mode

Strona 37

Changing Operating ModesSetting up your Monitor 3-57Setting up your Monitor Procedure 1. Enter the password by pressing the appropriate combination

Strona 38 - Parameter Modules

The Test Signals Function3-58 Setting up your Monitor Setting up your Monitor The Test Signals FunctionThe Test Signals function allows a check

Strona 39

The Test Signals FunctionSetting up your Monitor 3-59Setting up your Monitor Procedure 1. Press hardkey 2. Press softkey in the Instrument Confi

Strona 40

The Test Signals Function3-60 Setting up your Monitor Setting up your Monitor System Self-Test Values:Module Test Numeric Limits Test WaveformsE

Strona 41

Parameter Settings TransferSetting up your Monitor 3-61Setting up your Monitor Parameter Settings TransferParameter modules which are labeled on th

Strona 42 - Operating Levels

Parameter Settings Transfer3-62 Setting up your Monitor Setting up your Monitor or• by resetting all parameter settings to their default values

Strona 43

Parameter Settings Transfer MessagesSetting up your Monitor 3-63Setting up your Monitor Parameter Settings Transfer Messages The following messages

Strona 44

Parameter Settings Transfer Messages3-64 Setting up your Monitor Setting up your Monitor (Parameter name) reading settings from plug-in module&a

Strona 45 - Selection

Other Patients 4-14Other PatientsThis chapter describes the Other Patients function. It includes the following sections:• Overview. . . . . . . . .

Strona 46

Contents-2Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 47 - Getting into

Overview4-2 Other Patients Other Patients OverviewIf you are connected to a Philips patient care system the Other Patients function allows you t

Strona 48 - Filter/Mon/Diag

Philips Patient Care SystemOther Patients 4-3Other Patients Philips Patient Care SystemThe Philips Patient Care System is a network which transport

Strona 49

Philips Patient Care System4-4 Other Patients Other Patients The frame around the softkey for a bed in alarm will blink until the alarm is silen

Strona 50 - Panel Task

Philips Patient Care SystemOther Patients 4-5Other Patients The Alarm Bed Task Window shows the bed label of the other bed in alarm, the primary wa

Strona 51

Philips Patient Care System4-6 Other Patients Other Patients the alarm stack viewed by other monitors in the Other Patients group. No arrow next

Strona 52

Philips Patient Care SystemOther Patients 4-7Other Patients Configuring the Other Patients ControlsWith the Other Patients Controls you can place y

Strona 53 - Previous Item

Using Philips Patient Care System with an Arrhythmia Computer4-8 Other Patients Other Patients Using Philips Patient Care System with an Arrhyth

Strona 54 - Operation

Using Philips Patient Care System with an Arrhythmia ComputerOther Patients 4-9Other Patients If arrhythmia monitoring for your system is on, the h

Strona 55 - The CMS Computer Modules

Using Philips Patient Care System with an Arrhythmia Computer4-10 Other Patients Other Patients Arrhythmia Alarms on the 78560A and 78720 Arrhyt

Strona 56 - Maximum Voltage

Using Philips Patient Care System with an Arrhythmia ComputerOther Patients 4-11Other Patients R-on-T VPBs (**) For HR < 100, a VPB with R-R int

Strona 57

Contents-3ECG Output and Defibrillator Marker Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-34The V24 and V26 Parame

Strona 58 - Operating Rules to Remember

Using Philips Patient Care System with an Arrhythmia Computer4-12 Other Patients Other Patients Arrhythmia Alarms on the 78560A Arrhythmia Compu

Strona 59

Using Philips Patient Care System with an Arrhythmia ComputerOther Patients 4-13Other Patients Arrhythmia Alarms on the 3150A/3153A Arrhythmia Comp

Strona 60 - Using an ITE Display

Using Philips Patient Care System with an Arrhythmia Computer4-14 Other Patients Other Patients ** R-ON-T PVCs For HR <100, a PVC with R-R in

Strona 61

Extended Overview (CMS only)Other Patients 4-15Other Patients Extended Overview (CMS only)Note—The Extended Overview feature is optional and requir

Strona 62

Extended Overview (CMS only)4-16 Other Patients Other Patients To View an Extended Other Patients Bed1. Press .2. Depending on your monitor’s co

Strona 63 - Getting Started

Extended Overview (CMS only)Other Patients 4-17Other Patients 4. To view another bed in the group, select the appropriate group label.To select any

Strona 64 - 2-2 Getting Started

Extended Overview (CMS only)4-18 Other Patients Other Patients Note—If the gateway is configured to provide Alert Notification and the gateway i

Strona 65 - Getting Started 2-3

Alarm Functions 5-15Alarm FunctionsThis chapter provides general information on alarm functions and what to do if an alarm occurs. It includes the f

Strona 66 - 2-4 Getting Started

Alarm Display5-2 Alarm Functions Alarm Functions Alarm DisplayThe alarm messages are displayed across the top of the screen. When you switch the

Strona 67 - Getting Started 2-5

Alarm DisplayAlarm Functions 5-3Alarm Functions Alarm LampsRed and yellow alarm lamps on the control panel signify the severity of the cause of the

Strona 68 - Attaching the Patient

Contents-4Selecting the Number of Waves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13Procedur

Strona 69 - Contrast

Alarm Display5-4 Alarm Functions Alarm Functions If all the alarms have been turned off using the softkey or the hardkey, they can all be tu

Strona 70 - Messages

Alarm DisplayAlarm Functions 5-5Alarm Functions • Latching Alarms - The alarm continues when the alarm condition is eliminated. Latching alarms can

Strona 71 - Reserving a

Alarm Display5-6 Alarm Functions Alarm Functions The alarm reminder clock starts counting from the beginning again every time the key is press

Strona 72 - 2-10 Getting Started

Alarm DisplayAlarm Functions 5-7Alarm Functions Summary of Alarm Behavior Non-latching alarmsLatching alarms Visual Latching AudibleNon-Latching ha

Strona 73 - Getting Started 2-11

Alarm Display5-8 Alarm Functions Alarm Functions Alarm Priorities• Red alarms - Highest priority alarms, displayed in the top right corner of th

Strona 74

Alarm DisplayAlarm Functions 5-9Alarm Functions Viewing Alarm MessagesThe monitor alarms are arranged in increasing severity across the screen:1. P

Strona 75 - Setting up your Monitor

Alarm Display5-10 Alarm Functions Alarm Functions Individual Parameter AlarmsWhen you plug in a parameter module and connect the transducer or a

Strona 76 - Changing Display Screens

Alarm DisplayAlarm Functions 5-11Alarm Functions When an Alarm OccursThe alarm messages appear across the top of the screen. You need to identify t

Strona 77 - Selecting a Screen

Alarm Setup5-12 Alarm Functions Alarm Functions Alarm SetupIn the Alarm Limits Task Window you are able to review all of the alarm limits that h

Strona 78 - Freezing Waves (CMS only)

Alarm SetupAlarm Functions 5-13Alarm Functions Changing the Alarm Limits--> 1. Press to highlight the parameter that you want to adjust. 2. U

Strona 79 - What you Can Configure

Contents-5Changing the Configuration Set. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-53Procedu

Strona 80 - Changes to the Configuration

Alarm Setup5-14 Alarm Functions Alarm Functions Setting the Volume ControlThe QRS and the alarm tone volume can be set independently.1. Press h

Strona 81 - Setting up your Monitor 3-7

Alarm SetupAlarm Functions 5-15Alarm Functions The Nurse Call RelayNote—The nurse call relay is not available for the V24CT and V26CT.The nurse cal

Strona 82 - Display1 Setup

Alarm Setup5-16 Alarm Functions Alarm Functions

Strona 83

Recording Functions 6-16Recording FunctionsThis chapter contains information on the recordings that can be made using the patient monitors, and the

Strona 84

General Recorder Information6-2 Recording Functions Recording Functions General Recorder InformationRecorders The following recorders are availa

Strona 85

General Recorder InformationRecording Functions 6-3Recording Functions Controls and Indicators on the Plug-In Recordera. Continue light Flashes if

Strona 86

General Recorder Information6-4 Recording Functions Recording Functions Controls and Indicators on the 4-Channel Recorder (CMS only)a. Power On/

Strona 87 - Selecting the Number of Waves

General Recorder InformationRecording Functions 6-5Recording Functions Recorder CapabilitiesPlug-InRecorder(M1116A/B)Can record up to 3 waveforms (

Strona 88 - Changing the Wave Overlap

General Recorder Information6-6 Recording Functions Recording Functions CentralRecorder (1, 2or 4 channelrecorder)• Makes delayed recordings of

Strona 89

Types of RecordingsRecording Functions 6-7Recording Functions Types of RecordingsThe monitor produces several different types of recordings. Alarm

Strona 90

Contents-6Recorders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-

Strona 91 - Numerics

Delayed Recording6-8 Recording Functions Recording Functions Delayed RecordingDefinitions Delayed Recordings - recordings of waveforms that are

Strona 92

Delayed RecordingRecording Functions 6-9Recording Functions Configuring Delayed RecordingsFor Plug-Inand 4-ChannelBedsideRecorders 1. Select

Strona 93 - Additional

Delayed Recording6-10 Recording Functions Recording Functions For CentralRecorders → .1. Select the central recorder in the Delayed/Alarm Reco

Strona 94

Delayed RecordingRecording Functions 6-11Recording Functions Making Delayed RecordingsTo initiate a delayed recording on any recorder, press Delaye

Strona 95

Alarm Recording6-12 Recording Functions Recording Functions Alarm RecordingAlarm recordings are a type of delayed recording, that is, the record

Strona 96

Alarm RecordingRecording Functions 6-13Recording Functions Configuring Alarm Recordings 1. Press to highlight a parameter to change. 2. Pre

Strona 97

Alarm Recording6-14 Recording Functions Recording Functions Alarm Recording PrioritiesIn a delayed recording, waves are recorded in the order as

Strona 98

Alarm RecordingRecording Functions 6-15Recording Functions A yellow CO2 alarm would produce a recording ordered as:If a red P3 alarm occurred, the

Strona 99

Procedure Recordings6-16 Recording Functions Recording Functions Procedure RecordingsRecordings can be made during cardiac output measurements.C

Strona 100 - Notes on

Procedure RecordingsRecording Functions 6-17Recording Functions Making Procedure Recordings See Cardiac Output Section for details on cardiac outp

Strona 101

Contents-7Changing the Recorder Speed. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-43Changing the R

Strona 102 - CSA Display (CMS only)

Procedure Recordings6-18 Recording Functions Recording Functions → → See the Pressure Section for details on Pulmonary Artery Wedge Pressure

Strona 103

Procedure RecordingsRecording Functions 6-19Recording Functions (91MIA) 25 mm/secRef 01 DEC 8

Strona 104 - Wave Replace

Realtime Wave Recordings6-20 Recording Functions Recording Functions Realtime Wave RecordingsDefinitions Realtime Recordings - recordings of wav

Strona 105 - Trace Mode

Realtime Wave RecordingsRecording Functions 6-21Recording Functions Configuring Preset Recording Modes .1. Select the recorder and layout. a.

Strona 106 - Monitor Setup

Realtime Wave Recordings6-22 Recording Functions Recording Functions Making Preset Recordings if available , , or .Making Non-Preset Rec

Strona 107

Realtime Wave RecordingsRecording Functions 6-23Recording Functions Note—(CMS only): The number of recording channels available depends on the reco

Strona 108 - On/Off Setup

Realtime Wave Recordings6-24 Recording Functions Recording Functions For central recorders, the requests are run and queued according to the pri

Strona 109 - Adjusting the Volume Control

Realtime Vital Signs / Blood RecordingsRecording Functions 6-25Recording Functions Realtime Vital Signs / Blood RecordingsDefinitions 1. Realtime R

Strona 110 - 135/72 (94)

Realtime Vital Signs / Blood Recordings6-26 Recording Functions Recording Functions Depending on the number of parameters being monitored, a vit

Strona 111 - Adjusting the Date and Time

Realtime Vital Signs / Blood RecordingsRecording Functions 6-27Recording Functions Making a Single Vital Signs/Blood RecordingIf your monitor is co

Strona 112

ii NoticeThis document contains proprietary information which is protected by copyright. All Rights Reserved. Reproduction, adaptation, or tra

Strona 113

Contents-8Neonatal Event Review 9-1 Introduction to Neonatal Event Review. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strona 114 - Recordng Setup

Realtime Vital Signs / Blood Recordings6-28 Recording Functions Recording Functions If you are configured for vital signs / blood recordings and

Strona 115 - Other Patients Controls

Trended Vital Signs RecordingsRecording Functions 6-29Recording Functions Trended Vital Signs RecordingsYou can record the contents of the Vital Si

Strona 116 - The Status Log Function

Trended Vital Signs Recordings6-30 Recording Functions Recording Functions 6. Medical record numberTrend Data Depending upon the number of trend

Strona 117 - The Monitor Revision Function

Trended Vital Signs RecordingsRecording Functions 6-31Recording Functions 1. Time line. 7. Diastolic.2. Data that has been transferred. 8. Mean.3.

Strona 118

Trended Vital Signs Recordings6-32 Recording Functions Recording Functions Making Trended Vital Signs Recordings The trended data displayed in t

Strona 119 - Changing the Patient Category

Neonatal Event Review RecordingsRecording Functions 6-33Recording Functions Neonatal Event Review RecordingsTabular Neonatal Event RecordingsYou ca

Strona 120

Neonatal Event Review Recordings6-34 Recording Functions Recording Functions Depending upon the number of events stored, a neonatal event record

Strona 121 - Examples

Neonatal Event Review RecordingsRecording Functions 6-35Recording Functions OxyCRG Episode DataThe oxyCRG episode is recorded with the same labels,

Strona 122

oxyCRG Recordings6-36 Recording Functions Recording Functions oxyCRG RecordingsYou can record the contents of the oxyCRG display on an M1116B Pl

Strona 123

oxyCRG RecordingsRecording Functions 6-37Recording Functions The oxyCRG is shown in three distinct non-overlapping “channels” on the strip. The tim

Strona 124

Contents-9Explanation of Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11-12Maintenance Ch

Strona 125 - (HR) / Pulse

oxyCRG Recordings6-38 Recording Functions Recording Functions oxyCRG Alarm Recording Pressing the softkey allows you to choose between “S

Strona 126

Additional InformationRecording Functions 6-39Recording Functions Additional InformationAnnotations Delayed and realtime recordings are annotated w

Strona 127

Additional Information6-40 Recording Functions Recording Functions For example, the complete code (8AMIN360) at the beginning of a recording str

Strona 128 - Change ConfSet

Additional InformationRecording Functions 6-41Recording Functions CAL @@@ DEG C BY 5 SEC where @@@ is the signal value as follows:Cal Pulse @@@Scal

Strona 129 - Changing Operating Modes

Additional Information6-42 Recording Functions Recording Functions Changing the Recording LengthDelayed and realtime recordings can be extended

Strona 130 - Monitoring

Additional InformationRecording Functions 6-43Recording Functions Changing the Recorder SpeedThe recording speed can be changed for delayed, realti

Strona 131

Additional Information6-44 Recording Functions Recording Functions Inserting a Calibration Signal You can insert a calibration signal into the w

Strona 132 - The Test Signals Function

Additional InformationRecording Functions 6-45Recording Functions Layout Choices on Recorder M1117A Layout Sector 1 Sector 2 Sector 3 Sector 4#1125

Strona 133 - (CMS only)

Recording Status Messages6-46 Recording Functions Recording Functions Recording Status MessagesWhen a recording is running, you get a message in

Strona 134

Recording Status MessagesRecording Functions 6-47Recording Functions Note—Recording ink sometimes fades when covered with transparent tape. Please

Strona 135 - Parameter Settings Transfer

Contents-10Pressure Module Self-Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-16Performing

Strona 136

Accessories and Ordering Information6-48 Recording Functions Recording Functions Accessories and Ordering InformationYou can order parts and acc

Strona 137 - Silence Reset

Loading PaperRecording Functions 6-49Recording Functions Loading Paper Central RecordersTo load paper into the central recorders, see the operating

Strona 138

Loading Paper6-50 Recording Functions Recording Functions To Replace Paper in the Plug-In RecorderNote—To test if paper is loaded correctly, ini

Strona 139 - Other Patients

Loading PaperRecording Functions 6-51Recording Functions If the Paper isLoadedBackwardsIf you have loaded the paper backwards, to remove the roll,

Strona 140 - Overview

Loading Paper6-52 Recording Functions Recording Functions Cleaning the Print head in the Plug-In RecorderIf very slow speed (1 or 2 cm/min) reco

Strona 141 - Philips Patient Care System

Loading PaperRecording Functions 6-53Recording Functions message “no alarm recording available” will be displayed periodically on the monitor scree

Strona 142 - Patients

Loading Paper6-54 Recording Functions Recording Functions Note—No more than one sheet should be left hanging outside of the unit. Tear off extra

Strona 143

Loading PaperRecording Functions 6-55Note—The first sheet will come out folded, as shown.Cleaning the Roller on the Four Channel (M1117A) RecorderIf

Strona 144

Loading Paper6-56 Recording Functions

Strona 145 - Controls

Admit/Discharge/End Case 7-17Admit/Discharge/End CaseThis chapter describes the procedures for reviewing patient admit information, entering height

Strona 146 - Computer

The CMS and V24 and V26 Patient Monitors 1-11The CMS and V24 and V26 PatientMonitorsThis chapter provides an overview of the CMS Patient Monitoring

Strona 147 - Other Patients 4-9

Admitting a Patient7-2 Admit/Discharge/End Case Admit/Discharge/End Case Admitting a PatientThe patient monitors provide you with a display of d

Strona 148 - 4-10 Other Patients

Admitting a PatientAdmit/Discharge/End Case 7-3Admit/Discharge/End Case Admission Information (ICU mode) 1Neonatal onlyItemsRequired or Optional I

Strona 149 - Other Patients 4-11

Admitting a Patient7-4 Admit/Discharge/End Case Admit/Discharge/End Case Admission Information (OR Mode)Patient name will appear on the Main Scr

Strona 150 - 4-12 Other Patients

Admitting a PatientAdmit/Discharge/End Case 7-5Admit/Discharge/End Case name and medical record number at the bedside monitor (you can only change

Strona 151 - Other Patients 4-13

Admitting a Patient7-6 Admit/Discharge/End Case Admit/Discharge/End Case Moves the cursor up and downerases the input of a whole lineV24 and V26

Strona 152 - 4-14 Other Patients

Admitting a PatientAdmit/Discharge/End Case 7-7Admit/Discharge/End Case ICU Mode (Adult/Pediatric)ICU Mode (Neonatal)

Strona 153 - Extended Overview (CMS only)

Admitting a Patient7-8 Admit/Discharge/End Case Admit/Discharge/End Case OR Mode

Strona 154

Admitting a PatientAdmit/Discharge/End Case 7-9Admit/Discharge/End Case Discharging a Patient/Ending a CasePatient data for the current patient can

Strona 155 - ↑ preceding bed name)

Admitting a Patient7-10 Admit/Discharge/End Case Admit/Discharge/End Case endcase.tif

Strona 156 - Silence/Reset

Trends and Calculations 8-18Trends and CalculationsThis chapter gives you information on the patient data trends and calculations of the patient mon

Strona 157 - Alarm Functions

Introduction1-2 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors IntroductionThe Philips M1165/66/67/75/76/77 C

Strona 158 - Alarm Display

Introduction to Trends & Calculations8-2 Trends and Calculations Trends and Calculations Introduction to Trends & CalculationsThe patien

Strona 159 - Suspending

Viewing Patient DataTrends and Calculations 8-3Trends and Calculations Viewing Patient DataThere are two ways to view patient data:In either form,

Strona 160 - Resetting

Viewing Patient Data8-4 Trends and Calculations Trends and Calculations Extended Database (CMS only) If data is stored every minute, the values

Strona 161

Viewing Patient DataTrends and Calculations 8-5Trends and Calculations Vital Signs and Graph Trends are not functional during a transfer of data fr

Strona 162

Viewing Patient Data8-6 Trends and Calculations Trends and Calculations You can access the Blood Review task window by selecting from the Blood

Strona 163 - Alarm Functions 5-7

Viewing Patient DataTrends and Calculations 8-7Trends and Calculations Qualifying indicators (which have specific meaning) appear in the Blood Revi

Strona 164 - Priorities

Viewing Patient Data8-8 Trends and Calculations Trends and Calculations The table below describes how reference ranges appear on monochrome and

Strona 165

Viewing Patient DataTrends and Calculations 8-9Trends and Calculations Viewing Vital SignsPatient data can be shown in a table three minutes after

Strona 166 - Parameter

Viewing Patient Data8-10 Trends and Calculations Trends and Calculations For example, when you press from a Vital Signs display at one-minute

Strona 167 - When an

Viewing Patient DataTrends and Calculations 8-11Trends and Calculations Aperiodic parameters (parameters which are not monitored continuously such

Strona 168 - Alarm Setup

IntroductionThe CMS and V24 and V26 Patient Monitors 1-3The CMS and V24 and V26 Patient Monitors M1167/77A SystemDisplay Module M1095A 10.4” Flatsc

Strona 169 - Changing

Viewing Patient Data8-12 Trends and Calculations Trends and Calculations “Recording Functions” section (chapter 6) of this manual for further in

Strona 170 - Setting the

Viewing Patient DataTrends and Calculations 8-13Trends and Calculations Viewing Graph TrendsPatient data can be shown in a graph three minutes afte

Strona 171 - Call Relay

Viewing Patient Data8-14 Trends and Calculations Trends and Calculations The values plotted are the stored averaged values of readings taken eve

Strona 172

Viewing Patient DataTrends and Calculations 8-15Trends and Calculations If you change a value in Hemodynamics Calculations, (for example, you chang

Strona 173 - Recording Functions

Viewing Patient Data8-16 Trends and Calculations Trends and Calculations MarkingEventsThe Mark Event function enables you to mark events and the

Strona 174 - General Recorder Information

Viewing Patient DataTrends and Calculations 8-17Trends and Calculations Inserting aCalibrationMarkThe Mark Event function also enables you to inser

Strona 175 - In Recorder

Viewing Patient Data8-18 Trends and Calculations Trends and Calculations Procedure to Insert a Calibration Pulse When you press , you'

Strona 176 - 6-4 Recording Functions

Viewing Patient DataTrends and Calculations 8-19Trends and Calculations The calibration pulse varies according to the wave type:For pressure scales

Strona 177 - Capabilities

Performing and Reviewing Calculations8-20 Trends and Calculations Trends and Calculations Performing and Reviewing CalculationsThere are two cal

Strona 178 - 6-6 Recording Functions

Performing and Reviewing CalculationsTrends and Calculations 8-21Trends and Calculations Performing CalculationsCalculations can be performed using

Strona 179 - Types of Recordings

Introduction1-4 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors M1167/77A System with External Alarm DeviceDis

Strona 180 - Delayed Recording

Performing and Reviewing Calculations8-22 Trends and Calculations Trends and Calculations Calculation Task WindowCalculation Task Window for Tou

Strona 181 - Recordings

Performing and Reviewing CalculationsTrends and Calculations 8-23Trends and Calculations Notes—• If you plan to use monitored parameter values in C

Strona 182

Performing and Reviewing Calculations8-24 Trends and Calculations Trends and Calculations ResamplingVital SIgnsResampling of vital signs enables

Strona 183

Performing and Reviewing CalculationsTrends and Calculations 8-25Trends and Calculations Changing or Entering an Input ValueThe arrow keys are used

Strona 184 - Alarm Recording

Printing Reports8-26 Trends and Calculations Trends and Calculations Printing ReportsYou can print reports manually through task windows or auto

Strona 185

Printing ReportsTrends and Calculations 8-27Trends and Calculations To turn the scheduled report capability on or off, press in the Patient Data S

Strona 186 - Recording

Printing Reports8-28 Trends and Calculations Trends and Calculations Graph Trends ReportDAVID SCHULTZMR # : 012-34-56789DOB : 22 FEB 1945Bed

Strona 187 - Recording Functions 6-15

Printing ReportsTrends and Calculations 8-29Trends and Calculations Vital Signs ReportBlood Review ReportTable Report27 JAN 10:00 11:00 12:0

Strona 188 - Procedure Recordings

Printing Reports8-30 Trends and Calculations Trends and Calculations What to Do If Your Report Does Not PrintDuring normal report printing, mess

Strona 189 - Procedure

Printing ReportsTrends and Calculations 8-31Trends and Calculations MESSAGE: Report Request AcceptedMeaning: The report has been accepted for print

Strona 190 - softkey. If displayed

IntroductionThe CMS and V24 and V26 Patient Monitors 1-5The CMS and V24 and V26 Patient Monitors M1165/75A and M1166/76A SystemModel Types All syst

Strona 191 - 1 3/4 h x 5 3/8 w

Drug Calculator8-32 Trends and Calculations Trends and Calculations Drug CalculatorThe drug calculator provides drug infusion information at the

Strona 192 - Realtime Wave Recordings

Drug CalculatorTrends and Calculations 8-33Trends and Calculations By pressing , you will enter the Drug Calculation task window for the selected

Strona 193

Drug Calculator8-34 Trends and Calculations Trends and Calculations CautionIf you use the Drug Calculator for any patient other than the one cur

Strona 194 - Realtime Record

Drug CalculatorTrends and Calculations 8-35Trends and Calculations The Titration Table task window notes the relationship between dose and rate as

Strona 195 - Recording Functions 6-23

Drug Calculator8-36 Trends and Calculations Trends and Calculations

Strona 196

Neonatal Event Review 9-19Neonatal Event ReviewThis chapter contains information on how to view and interpret Neonatal Events and oxyCRG Episodes fo

Strona 197 - Recording Functions 6-25

Introduction to Neonatal Event Review9-2 Neonatal Event Review Neonatal Event Review Introduction to Neonatal Event ReviewThe Neonatal Event Re

Strona 198

Viewing Neonatal EventsNeonatal Event Review 9-3Neonatal Event Review Viewing Neonatal EventsOnce an event has been captured and stored by the moni

Strona 199 - VS/Blood PLUG-IN

Viewing Neonatal Events9-4 Neonatal Event Review Neonatal Event Review 4. The Manual Event ChannelThe graph provides an overview of the total nu

Strona 200

Viewing Neonatal EventsNeonatal Event Review 9-5Neonatal Event Review Graphical DetailsThe following information is displayed on the graph in the E

Strona 201 - Information

Introduction1-6 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors DisplayModulesBelow are labeled diagrams of th

Strona 202

Viewing Neonatal Events9-6 Neonatal Event Review Neonatal Event Review (1) Associated Event(2) Isolated Event(3)CursorThe severity of an event i

Strona 203 - Recording Functions 6-31

Viewing Neonatal EventsNeonatal Event Review 9-7Neonatal Event Review EventThreshold LineThe patient monitors can be adjusted in monitoring mode so

Strona 204 - Trends / Calcs

Viewing Neonatal Events9-8 Neonatal Event Review Neonatal Event Review Event Time The time at which the event was captured is stored with each e

Strona 205 - Neonatal

Viewing Neonatal EventsNeonatal Event Review 9-9Neonatal Event Review EventNumericsWhen the cursor is moved onto an event bar, specific numerical d

Strona 206 - Record Summary

Viewing Neonatal Events9-10 Neonatal Event Review Neonatal Event Review EventSummary LineThe event summary line is located above the graph, alon

Strona 207 - softkey

Viewing Neonatal EventsNeonatal Event Review 9-11Neonatal Event Review Example of an Event Summary LineThe above example of an event summary line,

Strona 208

Viewing Neonatal Events9-12 Neonatal Event Review Neonatal Event Review If the event storage is full, the newest (49th1) event will enter the gr

Strona 209

Viewing Neonatal EventsNeonatal Event Review 9-13Neonatal Event Review Note—The softkey is only active if at least one event is stored. Recording

Strona 210

Viewing oxyCRG Episodes9-14 Neonatal Event Review Neonatal Event Review Viewing oxyCRG EpisodesThe oxyCRG Episode task window displays a four mi

Strona 211 - Additional Information

Viewing oxyCRG EpisodesNeonatal Event Review 9-15Neonatal Event Review 1. The btb HR trend channel with a range from 60 up to 220 bpm.2. The SpO2 t

Strona 212

IntroductionThe CMS and V24 and V26 Patient Monitors 1-7The CMS and V24 and V26 Patient Monitors V24 and V26 Patient MonitorEach V24 and V26 Patien

Strona 213 - Recording Functions 6-41

Viewing oxyCRG Episodes9-16 Neonatal Event Review Neonatal Event Review The exact time an event was triggered is indicated by a vertical line wh

Strona 214

Viewing oxyCRG EpisodesNeonatal Event Review 9-17Neonatal Event Review Recording anEpisodeSelecting the softkey documents the selected oxyCRG epi

Strona 215

Adjusting Neonatal Event Review Settings9-18 Neonatal Event Review Neonatal Event Review Adjusting Neonatal Event Review SettingsThe Event Setup

Strona 216 - Calibration

Adjusting Neonatal Event Review SettingsNeonatal Event Review 9-19Neonatal Event Review The Event Setup task window enables you to adjust the follo

Strona 217 - Recording Functions 6-45

Adjusting Neonatal Event Review Settings9-20 Neonatal Event Review Neonatal Event Review BradycardiaEvent SettingsAlarm TriggeredThe Bradycardia

Strona 218 - Recording Status Messages

Adjusting Neonatal Event Review SettingsNeonatal Event Review 9-21Neonatal Event Review DesaturationEvent Settings Alarm TriggeredThe Desaturation

Strona 219

Adjusting Neonatal Event Review Settings9-22 Neonatal Event Review Neonatal Event Review Changing ItemContentsSelecting the softkey enables yo

Strona 220 - Description Part No

Data Transfer 10-110Data TransferThis chapter describes the Philips M1235A Data Transfer module. It includes the following sections:• Data Transfer

Strona 221 - Loading Paper

Data Transfer Module10-2 Data Transfer Data Transfer Data Transfer ModuleThe Philips M1235A Data Transfer module allows for rapid, convenient, a

Strona 222 - Plug-In

Data Transfer ModuleData Transfer 10-3Data Transfer !M1235ACTSDTMCMSCMSV24/V26

Strona 223 - Recording Functions 6-51

iiiThe M1165/66/75/76A Systems comply with UL544, CSA 22.2-125, IEC 601-1, EN 60601-1, and EN 60601-1-2 and carries Marking to Council Dire

Strona 224 - Print head in

Introduction1-8 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors Control PanelThe control panel consists of sof

Strona 225 - Recording Functions 6-53

Data Transfer Module10-4 Data Transfer Data Transfer Data transfer presents no interruptions to realtime monitoring and provides near-continuous

Strona 226

Data Transfer ModuleData Transfer 10-5Data Transfer When the module is unplugged, it retains data for at least one hour.After data is transferred t

Strona 227 - Recording Functions 6-55

Data Transfer Module10-6 Data Transfer Data Transfer The order in which the parameters are listed on the display is relative to the destination

Strona 228

Types of TransferData Transfer 10-7Data Transfer Types of TransferThere are two types of transfer to the module and one type of transfer to the mon

Strona 229 - Admit/Discharge/End Case

Types of Transfer10-8 Data Transfer Data Transfer Updates continue once per minute until the transfer is manually stopped.

Strona 230 - Admitting a Patient

Types of TransferData Transfer 10-9Data Transfer Append New Data to the Module (Patient not matching)Append New Data to the Module (Patient matchin

Strona 231

Types of Transfer10-10 Data Transfer Data Transfer To Monitor When a transfer is selected, all the module contents will be transferred to the

Strona 232

Types of TransferData Transfer 10-11Data Transfer Transferring Blood Analysis DataThe transfer of blood analysis data from one monitor to another w

Strona 233 - Patient

Types of Transfer10-12 Data Transfer Data Transfer Blood AnalysisData TransferScenarios• Transferring data from a source monitor to a destinatio

Strona 234 - V24 and V26 only

Types of TransferData Transfer 10-13Data Transfer • Transferring data from a source monitor via a transit monitor to a destination monitor requires

Strona 235 - ICU Mode (Neonatal)

IntroductionThe CMS and V24 and V26 Patient Monitors 1-9The CMS and V24 and V26 Patient Monitors Alarm Lamps The alarm lamps are lit when a red or

Strona 236

Combining Data10-14 Data Transfer Data Transfer Combining DataIn some situations, the patient name and/or ID in the monitor and module may not m

Strona 237

Combining DataData Transfer 10-15Data Transfer Database Conversion - Transfers to ModuleTransfers to the Monitor The following table shows the mapp

Strona 238

Combining Data10-16 Data Transfer Data Transfer 32 param, 12 hr, 1 min 16 param, 4 hr, 12 sec 32 param, 12 hr, 1 min 32 param, 12 hr, 1 min 32

Strona 239 - Trends and Calculations

Vital Signs, Blood Review and GraphsData Transfer 10-17Data Transfer Vital Signs, Blood Review and Graphs Some functions under Trends and Procedure

Strona 240 - Trends/Calcs

Vital Signs, Blood Review and Graphs10-18 Data Transfer Data Transfer The destination monitor creates the time stamp for transferred data by det

Strona 241 - Viewing Patient Data

TroubleshootingData Transfer 10-19Data Transfer TroubleshootingCautionIn a centrally networked system, if the previous patient is discharged only a

Strona 242

Performance Specifications10-20 Data Transfer Data Transfer Performance SpecificationsFor safety and environmental specifications, please refer

Strona 243 - Blood Mea

Monitor Installation and Patient Safety 11-111Monitor Installation and PatientSafetyThis chapter contains important information on the installation

Strona 244 - Blood Review

Introduction11-2 Monitor Installation and Patient Safety Monitor Installation and Patient Safety IntroductionThis section contains important inf

Strona 245 - Ranges/Units

IntroductionMonitor Installation and Patient Safety 11-3Monitor Installation and Patient Safety The M1205A V24 and V26 Patient Monitors and compati

Strona 246 - Vital Signs

Introduction1-10 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors The Handheld Keypad(CMS only):The handheld ke

Strona 247

Introduction11-4 Monitor Installation and Patient Safety Monitor Installation and Patient Safety CautionIf liquid is spilled on the equipment or

Strona 248 - Graph Trends

Installation InformationMonitor Installation and Patient Safety 11-5Monitor Installation and Patient Safety Installation InformationPower Source Re

Strona 249

Installation Information11-6 Monitor Installation and Patient Safety Monitor Installation and Patient Safety If the computer and display module

Strona 250 - Parameters

Installation InformationMonitor Installation and Patient Safety 11-7Monitor Installation and Patient Safety Safety Ground(ProtectiveEarth) (V24 and

Strona 251

Installation Information11-8 Monitor Installation and Patient Safety Monitor Installation and Patient Safety Combining EquipmentIf it is not evi

Strona 252 - softkey prior to . When

Installation InformationMonitor Installation and Patient Safety 11-9Monitor Installation and Patient Safety Environ-mentTo ensure a completely safe

Strona 253

Installation Information11-10 Monitor Installation and Patient Safety Monitor Installation and Patient Safety Philips M1167/77A CMS for Critical

Strona 254

Installation InformationMonitor Installation and Patient Safety 11-11Monitor Installation and Patient Safety Philips M1205A V24CT and V26CTCondensa

Strona 255 - Cal Signals

Installation Information11-12 Monitor Installation and Patient Safety Monitor Installation and Patient Safety Explanation of Symbols used Attent

Strona 256

Installation InformationMonitor Installation and Patient Safety 11-13Monitor Installation and Patient Safety Switch position is on/off.Contrast.Alt

Strona 257 - Mark Event

IntroductionThe CMS and V24 and V26 Patient Monitors 1-11The CMS and V24 and V26 Patient Monitors • To enter numbers and arithmetic symbols (labele

Strona 258

Maintenance Checks11-14 Monitor Installation and Patient Safety Monitor Installation and Patient Safety Maintenance ChecksBefore commencing moni

Strona 259 - Calculations

Maintenance ChecksMonitor Installation and Patient Safety 11-15Monitor Installation and Patient Safety This recommendation does not supersede any l

Strona 260 - Calculation Task Window

Maintenance Checks11-16 Monitor Installation and Patient Safety Monitor Installation and Patient Safety Patient Cables and LeadsInspect the pati

Strona 261 - Resample Vitals

Controls and ConnectorsMonitor Installation and Patient Safety 11-17Monitor Installation and Patient Safety Controls and ConnectorsThis section des

Strona 262 - On/Off DataTime

Controls and Connectors11-18 Monitor Installation and Patient Safety Monitor Installation and Patient Safety Connectors ofthe M1046AComputerModu

Strona 263

Controls and ConnectorsMonitor Installation and Patient Safety 11-19Monitor Installation and Patient Safety The Front Panel of the M1046B Computer

Strona 264 - Printing Reports

Controls and Connectors11-20 Monitor Installation and Patient Safety Monitor Installation and Patient Safety Connectors ofthe M1046BComputerModu

Strona 265

Controls and ConnectorsMonitor Installation and Patient Safety 11-21Monitor Installation and Patient Safety The Rear Panel of the M1046A/B Computer

Strona 266

Controls and Connectors11-22 Monitor Installation and Patient Safety Monitor Installation and Patient Safety Note—Do not apply a voltage of more

Strona 267 - Blood Review Report

Controls and ConnectorsMonitor Installation and Patient Safety 11-23Monitor Installation and Patient Safety (or receive inputs from the Ohmeda 7800

Strona 268 - Not Print

Introduction1-12 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors Hardkey Functions - press to silence an alarm

Strona 269

Controls and Connectors11-24 Monitor Installation and Patient Safety Monitor Installation and Patient Safety M1092A/M1094A Display Module.Maximu

Strona 270 - Drug Calculator

Controls and ConnectorsMonitor Installation and Patient Safety 11-25Monitor Installation and Patient Safety M1094B Display ModuleControls of theM10

Strona 271

Controls and Connectors11-26 Monitor Installation and Patient Safety Monitor Installation and Patient Safety 2. V AMP; this is the vertical ampl

Strona 272

Controls and ConnectorsMonitor Installation and Patient Safety 11-27Monitor Installation and Patient Safety 6. Human Interface Link Connector; this

Strona 273

Controls and Connectors11-28 Monitor Installation and Patient Safety Monitor Installation and Patient Safety M1095A Display ModuleControls of th

Strona 274

Controls and ConnectorsMonitor Installation and Patient Safety 11-29Monitor Installation and Patient Safety The Rear Panel of the M1109A External A

Strona 275 - Neonatal Event Review

Controls and Connectors11-30 Monitor Installation and Patient Safety Monitor Installation and Patient Safety loudspeaker is not covered and can

Strona 276

Controls and ConnectorsMonitor Installation and Patient Safety 11-31Monitor Installation and Patient Safety Caution• The System Power Outlet cannot

Strona 277 - Viewing Neonatal Events

Controls and Connectors11-32 Monitor Installation and Patient Safety Monitor Installation and Patient Safety M1167/77ASystemThe display module (

Strona 278 - Capture Episode

Controls and ConnectorsMonitor Installation and Patient Safety 11-33Monitor Installation and Patient Safety Lifting the Display ModuleThe M1092A di

Strona 279 - Graphical

IntroductionThe CMS and V24 and V26 Patient Monitors 1-13The CMS and V24 and V26 Patient Monitors The arrow keys consist of up/down/left/right keys

Strona 280

Controls and Connectors11-34 Monitor Installation and Patient Safety Monitor Installation and Patient Safety The V24 and V26 ConnectorsThe monit

Strona 281 - Neonatal Event Review 9-7

Controls and ConnectorsMonitor Installation and Patient Safety 11-35Monitor Installation and Patient Safety 4. Philips monitoring network connector

Strona 282

Controls and Connectors11-36 Monitor Installation and Patient Safety Monitor Installation and Patient Safety Assembling the V24 and V26The monit

Strona 283 - Neonatal Event Review 9-9

Accessories and Ordering InformationMonitor Installation and Patient Safety 11-37Monitor Installation and Patient Safety Accessories and Ordering I

Strona 284

Accessories and Ordering Information11-38 Monitor Installation and Patient Safety Monitor Installation and Patient Safety

Strona 285

Battery Information (V24CT and V26CT only) 12-112Battery Information (V24CT and V26CTonly)The V24CT and V26CT can be operated on AC line or DC batte

Strona 286 - Operating

AC and DC (Battery) Operation12-2 Battery Information (V24CT and V26CT only) Battery Information (V24CT and V26CT AC and DC (Battery) OperationYo

Strona 287

AC and DC (Battery) OperationBattery Information (V24CT and V26CT only) 12-3Battery Information (V24CT and V26CT Operating InstructionsThe following

Strona 288 - Viewing oxyCRG Episodes

AC and DC (Battery) Operation12-4 Battery Information (V24CT and V26CT only) Battery Information (V24CT and V26CT Note—When AC is connected and t

Strona 289 - Neonatal Event Review 9-15

AC and DC (Battery) OperationBattery Information (V24CT and V26CT only) 12-5Battery Information (V24CT and V26CT • The lead-acid batteries need not

Strona 290

V26CT/V24CT Power Supply1-14 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors V26CT/V24CT Power SupplyThe V24CT

Strona 291

Battery Indicator and Messages12-6 Battery Information (V24CT and V26CT only) Battery Information (V24CT and V26CT Battery Indicator and Messages

Strona 292 - Event Setup

Battery Indicator and MessagesBattery Information (V24CT and V26CT only) 12-7Battery Information (V24CT and V26CT • A low battery alarm INOP tone wi

Strona 293 - Criteria

External Battery Charger12-8 Battery Information (V24CT and V26CT only) Battery Information (V24CT and V26CT External Battery ChargerWe suggest t

Strona 294 - Change Content

Battery Care and MaintenanceBattery Information (V24CT and V26CT only) 12-9Battery Information (V24CT and V26CT Battery Care and MaintenanceThe Phil

Strona 295

Battery Care and Maintenance12-10 Battery Information (V24CT and V26CT only) Battery Information (V24CT and V26CT Care and Handling• Do not disas

Strona 296

Accessories and Ordering InformationBattery Information (V24CT and V26CT only) 12-11Battery Information (V24CT and V26CT Accessories and Ordering In

Strona 297 - Data Transfer

Accessories and Ordering Information12-12 Battery Information (V24CT and V26CT only) Battery Information (V24CT and V26CT

Strona 298 - Data Transfer Module

Maintenance 13-113MaintenanceThis chapter details the cleaning and maintenance procedures for the monitor. It includes the following sections:• Gene

Strona 299 - Data Transfer 10-3

General cleaning of the System13-2 Maintenance Maintenance General cleaning of the System The patient monitor should be kept free of dust and di

Strona 300 - MODULE MONITOR

General cleaning of the SystemMaintenance 13-3Maintenance 7. Do not allow cleaning agent to remain on any of the equipment surfaces - wipe it off i

Strona 301 - Transferred

V26CT/V24CT Power SupplyThe CMS and V24 and V26 Patient Monitors 1-15The CMS and V24 and V26 Patient Monitors Battery life includes the time during

Strona 302 - 10-6 Data Transfer

General Disinfecting of the System13-4 Maintenance Maintenance General Disinfecting of the SystemWe recommend that you disinfect the product onl

Strona 303 - Types of Transfer

General Disinfecting of the SystemMaintenance 13-5Maintenance 5. Do not pour liquid onto the system during cleaning.6. Never use abrasive material

Strona 304

Monitor Maintenance13-6 Maintenance Maintenance Monitor MaintenanceThe following is a list of the tasks and procedures to maintain the monitor.•

Strona 305

Monitor MaintenanceMaintenance 13-7Maintenance Preventative Maintenance Replace backlight tube assembly Cleaning procedures2.5 years of continu

Strona 306 - Data to Monitor

Monitor Maintenance13-8 Maintenance Maintenance Inspect the SystemThe following lists what to look for when inspecting the system:Inspect theExt

Strona 307 - → →

Monitor MaintenanceMaintenance 13-9Maintenance Perform a Start-up Sequence Test of the System The System automatically starts every time the system

Strona 308 - 10-12 Data Transfer

Monitor Maintenance13-10 Maintenance Maintenance Perform a System Self-TestThe following lists how to perform the system self-test.1. Examine al

Strona 309 - Collect New Data

Performance Assurance ChecksMaintenance 13-11Maintenance Performance Assurance Checks The following performance assurance checks are designed to be

Strona 310 - Combining Data

Performance Assurance Checks13-12 Maintenance Maintenance 4. If you are using the V24CT or V26CT, make sure the battery Charging and Charged LED

Strona 311 - Data Transfer 10-15

Performance Assurance ChecksMaintenance 13-13Maintenance Functional Testing ProceduresThe system has a module self-test which generates and display

Strona 312 - 10-16 Data Transfer

Parameter Modules1-16 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors Parameter Modules The parameter modules

Strona 313

Performance Assurance Checks13-14 Maintenance Maintenance ProcedureOverview1. Choose a patient category (Adult/Pedi/Neo) by pressing: 2. Dis

Strona 314 - 10-18 Data Transfer

Performance Assurance ChecksMaintenance 13-15Maintenance Performing the ECG Module and ECG/RESP Self-TestAuto Check - NoneSelf-Test The ECG and ECG

Strona 315 - Troubleshooting

Performance Assurance Checks13-16 Maintenance Maintenance Performing the Invasive Pressure Module Self-TestAuto Check - The check for these modu

Strona 316 - Performance Specifications

Performance Assurance ChecksMaintenance 13-17Maintenance Performing the NBP Module Self-TestAuto Check - The NBP module performs a check periodical

Strona 317

Performance Assurance Checks13-18 Maintenance Maintenance Performing the tcpO2/tcpCO2 Module Self-TestAuto Check - The check is performed automa

Strona 318

Performance Assurance ChecksMaintenance 13-19Maintenance Performing the CO2 Module Self-TestAuto Check - NoneSelf-Test - The self-test may be perfo

Strona 319

Performance Assurance Checks13-20 Maintenance Maintenance Performing the Blood Analysis Module Self-TestAuto Check - The Blood Analysis Module p

Strona 320 - Monitor Installation

Performance Assurance ChecksMaintenance 13-21Maintenance Tests for VueLink Module and Anesthetic Gas ModuleThere are no Self-Tests for the M1032A V

Strona 321 - Installation Information

Performance Assurance Checks13-22 Maintenance Maintenance

Strona 322

1Index of Volume 1AAC, 1-14operating instructions, 12-3accessoriestcpO2/tcpCO2, 6-48Adjusting, 2-7contrast, 2-7volume control, 3-35Admit, 7-2gene

Strona 323

Parameter ModulesThe CMS and V24 and V26 Patient Monitors 1-17The CMS and V24 and V26 Patient Monitors special Service Mode, either by your biomedi

Strona 324 - Equipment

2 Monitor Setup Selection Window, 3-2numeric display on/off, 3-17numeric positioning, 3-18operating modes, 3-55other patients controls, 3-41parameter

Strona 325 - Environ

3LLoading paperfour channel bedside recorder (M1117A), 6-53plug-in recorder (M1116A), 6-50plug-in recorder (M1116A/B), 6-49MMain Screen, 1-21Mon

Strona 326

4 inserting a calibration signal, 6-44loading paper, 6-49making calibrated ECG recordings, 6-23making delayed recordings, 6-11making non-preset record

Strona 327 - Condensa

iv Electromagnetic InterferenceAnomalies due to electromagnetic interference are not unique to the M1165/66/67/75/76/77A or the M1205A but are

Strona 328 - Explanation

Parameter Modules1-18 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors Parameter modules can be plugged into th

Strona 329

Parameter ModulesThe CMS and V24 and V26 Patient Monitors 1-19The CMS and V24 and V26 Patient Monitors For most types of parameter modules, the sys

Strona 330 - Maintenance Checks

Operating Levels1-20 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors Operating LevelsThere are three types of

Strona 331

Operating LevelsThe CMS and V24 and V26 Patient Monitors 1-21The CMS and V24 and V26 Patient Monitors Main Screen This display shows the waveforms

Strona 332 - Cables and

Operating Levels1-22 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors HR70135/72 (94)ST1ST2ABPPAP(23)SpO22ETCOI

Strona 333 - Controls and Connectors

Operating LevelsThe CMS and V24 and V26 Patient Monitors 1-23The CMS and V24 and V26 Patient Monitors Selection WindowThis is the first operating l

Strona 334 - Maximum Voltages

Operating Levels1-24 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors Task WindowThis is the second operating l

Strona 335 - Panel of the

Operating LevelsThe CMS and V24 and V26 Patient Monitors 1-25The CMS and V24 and V26 Patient Monitors Getting into the Operating LevelsBefore you c

Strona 336

Operating Levels1-26 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors There are two reasons for changing the se

Strona 337 - M1046A/B

Touch or Mouse/Trackball OperationThe CMS and V24 and V26 Patient Monitors 1-27The CMS and V24 and V26 Patient Monitors Touch or Mouse/Trackball Op

Strona 338 - ±12V to any of these outputs

Intended Use vIntended UseDescriptionThe Philips M1165/66/67/75/76/77A CMS Patient Monitoring System and the Philips M1205A V24 and V26 Patie

Strona 339 - Display

Touch or Mouse/Trackball Operation1-28 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors Main Screen The layout

Strona 340

Touch or Mouse/Trackball OperationThe CMS and V24 and V26 Patient Monitors 1-29The CMS and V24 and V26 Patient Monitors Control Panel Task WindowGe

Strona 341

Touch or Mouse/Trackball Operation1-30 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors Touch-ResponsiveObjects

Strona 342

Touch or Mouse/Trackball OperationThe CMS and V24 and V26 Patient Monitors 1-31The CMS and V24 and V26 Patient Monitors AlphanumericEntriesA Touchb

Strona 343

Touch or Mouse/Trackball Operation1-32 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors Disabling Touch/Mouse/T

Strona 344

The CMS Computer ModulesThe CMS and V24 and V26 Patient Monitors 1-33The CMS and V24 and V26 Patient Monitors The CMS Computer ModulesThe signals f

Strona 345

The CMS Computer Modules1-34 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors M1046B Computer ModuleThe M1046B

Strona 346

The V24 and V26 Parameter Module RackThe CMS and V24 and V26 Patient Monitors 1-35The CMS and V24 and V26 Patient Monitors The V24 and V26 Paramete

Strona 347 - Assembling

Operating Rules to Remember1-36 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors Operating Rules to Remember• P

Strona 348

Performance Specifications of the Philips DisplaysThe CMS and V24 and V26 Patient Monitors 1-37The CMS and V24 and V26 Patient Monitors Performance

Strona 349 - Note—The FIO

Indications for Usevi Indications for UseConditionThe Philips M1165/66/67/75/76/77A CMS Patient Monitoring System and the Philips M1205A V24 a

Strona 350 - Connectors

Using an ITE Display1-38 The CMS and V24 and V26 Patient Monitors The CMS and V24 and V26 Patient Monitors Using an ITE DisplayThe M1167A/77A CM

Strona 351

Using an ITE DisplayThe CMS and V24 and V26 Patient Monitors 1-39EMC 1. The display must either fulfill the requirements of the EN60601-1-2 {IEC 601

Strona 352

Using an ITE Display1-40 The CMS and V24 and V26 Patient Monitors

Strona 353

Getting Started 2-12Getting StartedNow that you have been introduced to the monitor you are probably ready to start using it. This chapter will help

Strona 354

Setting up the Monitor (V24 and V26 only)2-2 Getting Started Getting Started Setting up the Monitor (V24 and V26 only)1. Attach the parameter mo

Strona 355 - (V24CT and V26CT

Setting up the Monitor (V24 and V26 only)Getting Started 2-3Getting Started 2. If you are using the V24CT and V26CT and the batteries are not alrea

Strona 356 - AC and DC (Battery) Operation

Setting up the Monitor (V24 and V26 only)2-4 Getting Started Getting Started 4. Switch on the system using the power On\Off switch. A self test

Strona 357 - Instructions

Setting up the Parameter ModulesGetting Started 2-5Getting Started Setting up the Parameter Modules1. Decide which parameters you want to monitor.

Strona 358

Attaching the Patient2-6 Getting Started Getting Started Attaching the Patient1. Attach the electrodes, probes, transducers, and insert pressure

Strona 359

Attaching the PatientGetting Started 2-7Getting Started Adjusting Screen ContrastV24 and V26:The brightness or contrast of the screen display can b

Strona 360 - 30/17 (23)

Indications for Use viiPrescription Versus Over-the-CounterThe Philips M1165/66/67/75/76/77A CMS Patient Monitoring System and the Philips M1

Strona 361

Attaching the Patient2-8 Getting Started Getting Started Starting MonitoringThe patient's vital signs are now appearing on the screen. The

Strona 362 - External Battery Charger

Attaching the PatientGetting Started 2-9Getting Started this message stays as long as the ECG module is not plugged in. Prompt and status messages

Strona 363 - Battery Care and Maintenance

Attaching the Patient2-10 Getting Started Getting Started Power FailureIf a power failure occurs, any alarm conditions that were stored in the m

Strona 364 - Handling

Attaching the PatientGetting Started 2-11Getting Started If your monitor is connected to a Philips patient care system while the monitor is in stan

Strona 365 - M1278A Battery Charger

Attaching the Patient2-12 Getting Started Getting Started

Strona 366

Setting up your Monitor 3-13Setting up your MonitorThis chapter describes the characteristics of your system that can be changed during monitoring,

Strona 367 - Maintenance

Changing Display Screens3-2 Setting up your Monitor Setting up your Monitor Changing Display Screens A display screen is a pre-selected arrangem

Strona 368 - 13-2 Maintenance

Selecting a ScreenSetting up your Monitor 3-3Setting up your Monitor Selecting a ScreenIf the screen you want to change is not the currently active

Strona 369 - Maintenance 13-3

Freezing Waves (CMS only)3-4 Setting up your Monitor Setting up your Monitor Freezing Waves (CMS only)You can freeze any wave movement on the sc

Strona 370 - 13-4 Maintenance

What you Can ConfigureSetting up your Monitor 3-5Setting up your Monitor What you Can ConfigureChanges and adjustments you make to the settings in

Strona 371 - Maintenance 13-5

Indications for Useviii Warnings, Cautions, and Notes Warnings, cautions, and notes are used throughout this User's Manual to give you ad

Strona 372 - Monitor Maintenance

Changes to the Configuration3-6 Setting up your Monitor Setting up your Monitor Changes to the ConfigurationTo make changes to the settings, you

Strona 373 - Maintenance 13-7

Making Changes to the Main DisplaySetting up your Monitor 3-7Setting up your Monitor Making Changes to the Main DisplayThe following describes how

Strona 374 - Inspect the

Assigning Waves to Screen Channels3-8 Setting up your Monitor Setting up your Monitor Assigning Waves to Screen ChannelsYou can assign realtime

Strona 375 - Maintenance 13-9

Assigning Waves to Screen ChannelsSetting up your Monitor 3-9Setting up your Monitor Procedure 1. Press repeatedly to select a channel on the scr

Strona 376 - System Self

Selecting a Screen3-10 Setting up your Monitor Setting up your Monitor Selecting a ScreenYou can switch between the five available screens (A-E)

Strona 377 - Performance Assurance Checks

Selecting Screen Labels for Realtime Display ScreensSetting up your Monitor 3-11Setting up your Monitor Selecting Screen Labels for Realtime Displa

Strona 378 - 13-12 Maintenance

Selecting Screen Labels for Realtime Display Screens3-12 Setting up your Monitor Setting up your Monitor h. Inductni. Non Invasivej. OxyCRG3.k.

Strona 379

Selecting the Number of WavesSetting up your Monitor 3-13Setting up your Monitor Selecting the Number of WavesYou can select how many waves you wan

Strona 380

Changing the Wave Overlap3-14 Setting up your Monitor Setting up your Monitor Changing the Wave OverlapThe wave overlap refers to the position o

Strona 381 - Self-Test

Selecting Realtime Wave SpeedsSetting up your Monitor 3-15Setting up your Monitor Selecting Realtime Wave SpeedsThe same speed can be selected for

Strona 382 - Module Self

Indications for Use ix• Exposure of electrical contacts or connections to saline or other liquids and gels is dangerous. Electrical contacts

Strona 383 - Refernce

Selecting Realtime Wave Speeds3-16 Setting up your Monitor Setting up your Monitor Press the softkey to return to the Display Setup Window, pr

Strona 384 - Test Signals

Selecting Realtime Wave SpeedsSetting up your Monitor 3-17Setting up your Monitor Numerics On/OffTo tailor the numeric display format to specific n

Strona 385

Selecting Realtime Wave Speeds3-18 Setting up your Monitor Setting up your Monitor If numerics are switched off, the status message “Not all num

Strona 386 - 13-20 Maintenance

Selecting Realtime Wave SpeedsSetting up your Monitor 3-19Setting up your Monitor Numeric Sizes:Parameter numerics on the screen are displayed in 2

Strona 387 - Gas Module

Selecting an Application Window3-20 Setting up your Monitor Setting up your Monitor Selecting an Application WindowOne of three application wind

Strona 388

Selecting an Application WindowSetting up your Monitor 3-21Setting up your Monitor Note—If you have configured a second or third independent displa

Strona 389 - Index of Volume 1

Displaying Split Screen Trends3-22 Setting up your Monitor Setting up your Monitor Displaying Split Screen Trends Split Screen trends display th

Strona 390

Displaying Split Screen TrendsSetting up your Monitor 3-23Setting up your Monitor The alarm limits are represented by a rectangle, filled with the

Strona 391

Displaying Split Screen Trends3-24 Setting up your Monitor Setting up your Monitor In addition to this, the trend to which the alarm limits appl

Strona 392

oxyCRG DisplaySetting up your Monitor 3-25Setting up your Monitor oxyCRG DisplayThe oxyCRG (oxygen CardioRespiroGram) display provides a simultaneo