35 Default Settings Appendix Respiration Default Settings
374
Respiration Default Settings
SpO
2
Default Settings
Respiration Settings
Factory defaults H10 / H20 / H40 H30 (deviations from H10/20/40)
Adult Pedi Neo Adult Pedi Neo
High Limit 30 rpm 100 rpm
Low Limit 8 rpm 30 rpm
Apnea Time 20 sec
Alarms On
Resp On Off
Auto/Manual Auto (Trigger Mode)
Resp Speed 6.25mm/s
Color Yellow White
SpO
2
Settings Factory Adult Factory Pedi Factory Neo
Alarms On
QRS Volume 1
Ton e Modul at io n Ye s
Ton e Mod. Ty pe E nh a nce d
Speed 25mm/s
Perfusion On
Average 10 sec
NBP Alarm Suppr. On
Extd. Auto OnOff Disabled
Color Cyan
SpO
2
Alarm Default Settings
Setting adult pediatric neonatal
Desat Limit 80 80 80
Low Limit 90 90 85
High Limit 100 100 95
Desat delay 20 sec 20 sec 20 sec
High Alarm delay 10 sec 10 sec 10 sec
Low Alarm delay 10 sec 10 sec 10 sec
Parameter Alarms On/Off on on on
Label SpO
2
SpO
2
SpO
2
Pulse Settings
Pulse (SpO2) On/Off on on on