23 Event Surveillance Viewing Events
248
Counting Combi-Events
If one or more events occur during the same Episode Time, the monitor combines them and displays
them as distinct events in one event episode, called a combi-event. The first event is the trigger event,
and the others are follow-up events. For example, if an apnea event is followed 40 seconds later by a
brady event, the brady event is not counted as a single event but as part of the apnea event.
Counting Neonatal Event Review (NER) Events
For neonatal events (NER, formerly “OxyCRG”), apnea events (A), bradycardia events (B), and
combinations of these events are counted and classified by the event counter in the Event Summary. If
they are associated with a Desaturation (
D), this is also marked. Manual events (M) are counted
separately. In the example below,
A(D):2(1) indicates that two apnea events occurred and one of
them was associated with a desaturation.
16 Stored Events: A(D):2(1) B(D):7(1) AB (D): 1(1) D:6 M:0